Scanning Non-engaging Locators with a Reference Jig

Prerequisite

The Setting-up the Scanner for Implant Overdenture Scanning step must be completed.

Steps

  1. From the Scan List , click Implant locators.
  2. In the Camera View, align the blue numbered circles over their corresponding non-engaging locators.
    Note:

    Numbered circles must each fully cover their locator. Set a numbered circle over each implant, even if you did not have enough locators.

  3. Right-click the numbered circles that do not have any locator, this will disable them i.e. they will not be scanned.
    Note:

    Enabled circles are blue. Disabled circles are transparent.

  4. Align the three remaining blue circles over the spheres of the Reference Jig.
    Note:

    Use the spheres on a same row.

    Note:

    If some spheres are outside of the Camera View, gently push the scanning table forward until all spheres are displayed in the Camera View. This does not affect the selected scanning area as the camera acts like a targeting system and tracks the location of the circles already scanned.

  5. Click Scan.
    Note:

    Scanning can take several minutes.

  6. Follow displayed instructions.
  7. If you had enough non-engaging locators for all implants go to the next step otherwise, continue to sub-step a.
    Note:

    It is crucial that the model and table are not moved at all during the following sub-steps in order to maintain the same spatial relationship between the different scan groups of locators.

    1. Right-click the disabled circles to enable them.
    2. Remove the non-engaging locators from the positions already scanned, and place them on the implants that did not have any locators.
    3. Click Scan.
    4. Click No otherwise all of your previous position locator scans will be deleted.
    5. Follow displayed instructions.

Next step

Scanning the Model