Steps

- 
                Screw the calibration ball into the center hole of the scanning table.
                
Each scanner is delivered with its own calibration sphere. Do
                        not mix spheres of different equipment. Calibration spheres from 2G and 1G
                        scanners are slightly different and are not
                        interchangeable.
             
- 
                1G scanner
                    specific tasks : adjust the scan table.
                
                    - 
                        Set the scanning table to the lowest position by rotating the Height
                            Adjustment screw counterclockwise .
                    
 
                    - 
                        Using the Level Adjustment screw, level the scanning table . 
                    
 
                
             
- 
                From the Maintenance panel in
                    the scanner module, click Calibrate.
                
                    Note: Calibration can take up to 30 or 45 minutes. A message will be displayed
                        indicating that calibration is completed.
                 
             
- 
                Confirm calibration.
            
 
        Result
            Scanner is ready to be used.