Check calibration scan

To quickly check whether a full calibration scan is needed, perform a Check calibration scan.

In the Scanner calibration dialog, click Check calibration.

Place the calibration ball in the scanner and click Start.

A progress window is shown.

To stop the check calibration scan, click Abort. A warning message appears, asking for your confirmation.

If the calibration was successful, you will get following Calibration check successful message.

Click Finish.

Should the calibration procedure fail, perform a full calibration scan. A full calibration scan will be necessary before performing any scans. If the failure persists, please contact your local customer support.